Product Overview
The HA1600 Radioactive Aerosol & Iodine Monitor uses a vacuum pump to extract the air in the target area, collects the aerosol and iodine in the air in the target area through filter paper and iodine box adsorption, and then measures the radioactivity of the collected filter paper and iodine box to analyze the radioactivity level of aerosol particles and free iodine in the air.
Product Overview
The HA1600 Radioactive Aerosol & Iodine Monitor uses a vacuum pump to extract the air in the target area, collects the aerosol and iodine in the air in the target area through filter paper and iodine box adsorption, and then measures the radioactivity of the collected filter paper and iodine box to analyze the radioactivity level of aerosol particles and free iodine in the air.
Features
① Simultaneously measure the radioactivity levels of aerosols and iodine in the air;
② Built-in medical nuclide library, and support custom addition;
③ The lead chamber shielding factor is >95%, reducing environmental interference;
④ Standard process measurement, storing detailed energy spectrum and process records, matching regulatory specifications;
⑤ Adopting stripping technology, sampling deposition correction, decay correction, environmental radon thorium correction and other technologies to ensure detection accuracy;
⑥ Real-time monitoring of air pump operation, immediate stop of the pump when sampling is abnormal, improving the safety of equipment use.
Features
① Simultaneously measure the radioactivity levels of aerosols and iodine in the air;
② Built-in medical nuclide library, and support custom addition;
③ The lead chamber shielding factor is >95%, reducing environmental interference;
④ Standard process measurement, storing detailed energy spectrum and process records, matching regulatory specifications;
⑤ Adopting stripping technology, sampling deposition correction, decay correction, environmental radon thorium correction and other technologies to ensure detection accuracy;
⑥ Real-time monitoring of air pump operation, immediate stop of the pump when sampling is abnormal, improving the safety of equipment use.
Technical Parameters
Product Model |
HA1600 |
Detector Type |
NaI(Tl) crystal |
Detector size |
Φ50×50mm |
Detection nuclide type |
67Ga, 125I, 131I, 99mTc, 201Tl, single photon nuclides (nuclides can be edited and added) |
Lead room shielding factor |
>95% |
Effective sampling flow rate |
0~50LPM |
Measuring range |
3.7~3.7×106Bq/m3 |
Energy resolution |
≤7.5%(662keV,137Cs) |
MCA Channels |
2048 |
Detection efficiency |
131I total absorption peak 3.49% (131I is evenly distributed in the iodine box) |
power |
≤250w |
Technical Parameters
Product Model |
HA1600 |
Detector Type |
NaI(Tl) crystal |
Detector size |
Φ50×50mm |
Detection nuclide type |
67Ga, 125I, 131I, 99mTc, 201Tl, single photon nuclides (nuclides can be edited and added) |
Lead room shielding factor |
>95% |
Effective sampling flow rate |
0~50LPM |
Measuring range |
3.7~3.7×106Bq/m3 |
Energy resolution |
≤7.5%(662keV,137Cs) |
MCA Channels |
2048 |
Detection efficiency |
131I total absorption peak 3.49% (131I is evenly distributed in the iodine box) |
power |
≤250w |