Product Overview
HA4201 Food and Water Radioactivity Detector is mainly used for rapid detection of water samples, food samples, and biological samples. The instrument uses NaI (Tl) crystal detector, which has the characteristics of high sensitivity, fast response, integrated structure, touch color display, and is easy to use.
Product Overview
HA4201 Food and Water Radioactivity Detector is mainly used for rapid detection of water samples, food samples, and biological samples. The instrument uses NaI (Tl) crystal detector, which has the characteristics of high sensitivity, fast response, integrated structure, touch color display, and is easy to use.
Features
① Rapid analysis: can quickly measure the activity of γ-radioactive nuclides such as 60Co, 137Cs, and 131I in samples;
② Ultra-high automation: integrates control, data acquisition, data analysis and other functions;
③ Support and expansion computer control: provides external interface, configures PC operating software, supports data transmission, storage, and printing;
④ Powerful functions: background subtraction, automatic peak search, automatic spectrum stabilization, automatic radioactivity measurement report, and custom radionuclide library;
⑤ Ultra-long battery life: built-in large-capacity lithium battery, can be used in the laboratory or carried out.
Features
① Rapid analysis: can quickly measure the activity of γ-radioactive nuclides such as 60Co, 137Cs, and 131I in samples;
② Ultra-high automation: integrates control, data acquisition, data analysis and other functions;
③ Support and expansion computer control: provides external interface, configures PC operating software, supports data transmission, storage, and printing;
④ Powerful functions: background subtraction, automatic peak search, automatic spectrum stabilization, automatic radioactivity measurement report, and custom radionuclide library;
⑤ Ultra-long battery life: built-in large-capacity lithium battery, can be used in the laboratory or carried out.
Technical Parameters
Product Model |
HA4201 |
Detector Type |
NaI(Tl) crystal |
Detector size |
40×40mm |
Measuring range |
10~105Bq/L |
Energy range |
30keV~3MeV (10MeV optional) |
Energy resolution |
≤7.5%(@662keV,137Cs) |
Energy Linearity |
≤1% |
MCA Channels |
512、1024、2048 |
Measurement uncertainty |
<10%(Sample radioactivity>37Bq/kg) |
Timing Measurement |
1~86400s continuously adjustable |
Background Shielding |
Thickness 20mm 4πSr lead shield |
Detection nuclide type |
Can detect 226Ra, 232Th, 40K, 222Rn, 137Cs, 241Am and other radioactive nuclides, the nuclide library can be edited |
Power supply |
DC12V, built-in rechargeable lithium battery |
Operating temperature |
0℃~40℃ |
Technical Parameters
Product Model |
HA4201 |
Detector Type |
NaI(Tl) crystal |
Detector size |
40×40mm |
Measuring range |
10~105Bq/L |
Energy range |
30keV~3MeV (10MeV optional) |
Energy resolution |
≤7.5%(@662keV,137Cs) |
Energy Linearity |
≤1% |
MCA Channels |
512、1024、2048 |
Measurement uncertainty |
<10%(Sample radioactivity>37Bq/kg) |
Timing Measurement |
1~86400s continuously adjustable |
Background Shielding |
Thickness 20mm 4πSr lead shield |
Detection nuclide type |
Can detect 226Ra, 232Th, 40K, 222Rn, 137Cs, 241Am and other radioactive nuclides, the nuclide library can be edited |
Power supply |
DC12V, built-in rechargeable lithium battery |
Operating temperature |
0℃~40℃ |