Product Overview
HA-BASE digital multi-channel analyzer adopts large-scale integrated circuits and FPGA control logic, with low power consumption and high performance; it integrates high voltage regulation, digital signal processing and other functions, and is suitable for a variety of spectrum analysis from low count rate to high count rate. HA-BASE digital multi-channel analyzer can be directly connected to scintillation detectors such as NaI, CsI, LaBr3, etc., and realizes energy spectrum analysis function through multi-channel energy spectrum analysis software.
Product Overview
HA-BASE digital multi-channel analyzer adopts large-scale integrated circuits and FPGA control logic, with low power consumption and high performance; it integrates high voltage regulation, digital signal processing and other functions, and is suitable for a variety of spectrum analysis from low count rate to high count rate. HA-BASE digital multi-channel analyzer can be directly connected to scintillation detectors such as NaI, CsI, LaBr3, etc., and realizes energy spectrum analysis function through multi-channel energy spectrum analysis software.
Features
① Plug and Play: USB interface is used, the maximum throughput rate can reach 12Mbps, and it can be plug and play with laptops or PCs, no external power supply is required, and the transmission rate is fast;
② High measurement accuracy: It has an automatic spectrum stabilization mechanism, which can automatically adjust the software/hardware gain according to the peak position change, effectively avoiding the peak drift caused by environmental changes (such as temperature difference) and ensuring measurement accuracy;
③ Strong applicability: It adopts high-speed digital ADC circuit design, and can adjust the optimal sampling frequency according to the pulse signal width of the detector, which has strong applicability;
④ Low power consumption: The power consumption of the whole machine is less than 2w;
⑤ Positive and negative pulses are optional: The input pulse polarity is optional, and the analysis of positive pulse or negative pulse input can be selected according to actual needs.
Features
① Plug and Play: USB interface is used, the maximum throughput rate can reach 12Mbps, and it can be plug and play with laptops or PCs, no external power supply is required, and the transmission rate is fast;
② High measurement accuracy: It has an automatic spectrum stabilization mechanism, which can automatically adjust the software/hardware gain according to the peak position change, effectively avoiding the peak drift caused by environmental changes (such as temperature difference) and ensuring measurement accuracy;
③ Strong applicability: It adopts high-speed digital ADC circuit design, and can adjust the optimal sampling frequency according to the pulse signal width of the detector, which has strong applicability;
④ Low power consumption: The power consumption of the whole machine is less than 2w;
⑤ Positive and negative pulses are optional: The input pulse polarity is optional, and the analysis of positive pulse or negative pulse input can be selected according to actual needs.
Technical Parameters
Product Model |
HAS-BASE |
Gain adjustment range |
0.2~1.5 |
Input pulse |
Both positive and negative pulses are possible |
ADC Clock |
12M, 24M, 36M, 48M, 60M (optional) |
ADC resolution |
14-bit (214) |
Time threshold |
0~2100000s |
Maximum measurable signal pulse amplitude |
±2.0V |
MCA Channels |
256、512、1024、2048 |
Maximum number of quantization levels |
±2048 |
High pressure range |
0~1000V |
Linear |
Integral nonlinearity ±0.05%; differential nonlinearity ±1.0% |
stability |
Long-term stable and reliable operation, 72-hour zero drift is less than 1 channel, Gain instability is less than 0.2% in 72 hours |
Temperature coefficient |
Zero temperature coefficient is less than 0.2 °C Gain temperature coefficient is less than 0.1%/℃ |
Communication interface power consumption |
≤2w |
Technical Parameters
Product Model |
HAS-BASE |
Gain adjustment range |
0.2~1.5 |
Input pulse |
Both positive and negative pulses are possible |
ADC Clock |
12M, 24M, 36M, 48M, 60M (optional) |
ADC resolution |
14-bit (214) |
Time threshold |
0~2100000s |
Maximum measurable signal pulse amplitude |
±2.0V |
MCA Channels |
256、512、1024、2048 |
Maximum number of quantization levels |
±2048 |
High pressure range |
0~1000V |
Linear |
Integral nonlinearity ±0.05%; differential nonlinearity ±1.0% |
stability |
Long-term stable and reliable operation, 72-hour zero drift is less than 1 channel, Gain instability is less than 0.2% in 72 hours |
Temperature coefficient |
Zero temperature coefficient is less than 0.2 °C Gain temperature coefficient is less than 0.1%/℃ |
Communication interface power consumption |
≤2w |